pH Range
|
-2.00 to 16.00 pH
|
Resolution
|
0.01 pH
|
Accuracy
|
± 0.01 pH
|
Calibration Points
|
Up to 5 (USA & NIST) / Up to 6 (DIN)
|
Buffer Options
|
USA, NIST, DIN
|
ORP Range
|
± 2000 mV
|
Resolution
|
0.1 mV (< ±1000 mV), 1 mV (≥ ±1000 mV)
|
Accuracy
|
±0.3 mV (< ±1000 mV), 0.3% of reading (≥ ±1000 mV)
|
Calibration Option
|
Yes
|
DO Range
|
0.00 to 20.00 mg/L / 0.0 to 200.0 %
|
Resolution
|
0.01 mg/L or 0.1 %
|
Accuracy
|
±0.1 mg/L
|
Salinity Compensation
|
0.0 to 40.0 ppt
|
Barometric Pressure Compensation
|
Yes
|
Temperature Compensation
|
Yes
|
Calibration Points
|
Up to 2
|
DO Probe Type
|
Galvanic integrated with temperature sensor
|
Temperature Range
|
-30.0 to 130.0 °C / -22.0 to 266.0 °F
|
Resolution
|
0.1 °C / °F
|
Accuracy
|
± 0.5 °C / ± 0.9 °F
|
Calibration Option
|
Yes
|
Memory Data
|
500
|
Auto Log Data
|
Yes
|
Measurement Modes
|
Auto Stable, Auto Hold, Real Time
|
Offset & Average Slope Display
|
Yes
|
Calibration Alarm
|
Yes (Programmable: up to 90 days)
|
Auto Shut-Off
|
Yes (Programmable: up to 30 minutes)
|
Electrode Status
|
Yes (On screen display)
|
Diagnostic Messages
|
Yes
|
Software Upgrade
|
Yes via PC (USB) cable PN 3200779639
|
Meter Inputs
|
2 x BNC (for pH and DO electrodes), 2 x phono (for temperature sensors)
|
Meter Output
|
phono (for software upgrade only)
|
Display
|
Custom LCD with backlight
|
Housing
|
IP67, shock & scratch resistant, non-slip
|
Power Requirement
|
2 x AA batteries
|
Battery Life
|
> 500 hours
|
Meter Dimensions
|
160 (L) x 80 (W) x 40.60 (H) mm
|
Meter Weight
|
Approx. 260g (with batteries) / 216g (without batteries)
|